Measures
|
Dissolved oxygen, °C
|
|
% saturation Range
|
0.00 to 200%
|
|
% saturation Resolution
|
1%
|
|
% saturation Accuracy
|
±1.5% full-scale
|
|
mg/l (ppm) Range
|
0.0 to 20.00 mg/l (ppm)
|
|
mg/l (ppm) Resolution
|
0.01 mg/l (ppm)
|
|
mg/l (ppm) Accuracy
|
±1.5% full-scale
|
|
Temperature Range
|
0 to 50°C
|
|
Temperature Resolution
|
0.1°C
|
|
Temperature Accuracy
|
±0.5°C
|
|
Calibration Dissolved Oxygen
|
one or two points (100% and 0% saturation)
|
|
Calibration Temperature
|
offset up to ±5°C
|
|
Control Type
|
on/off or proportional (selectable)
|
|
Number of Inputs
|
one
|
|
Number of Set Points
|
two (high and low)
|
|
Separate Alarm Relay
|
yes
|
|
Output Control
|
2 SPDT relays, 6 A at 110 VAC, 250 VAC maximum
|
|
Output Alarm
|
1 SPDT relay, 6 A at 110 VAC, 250 VAC maximum
|
|
Output Pulse
|
pulse length (0.5 to 20 sec) or pulse frequency (60 to 120 pulses per min)
|
|
Output Current
|
isolated 0-20.4-20 mA, with 20% to 100% full-scale scalable boundaries, 600 maximum load
|
|
Hysteresis
|
0.1 to 1.0 mg/l; 1 to 10%
|
|
Relay Selay
|
selectable, 0 to 2000 seconds
|
|
Input Impedance
|
10^12 M
|
|
Electrical Isolation
|
yes, galvanically
|
|
Salinity Correction
|
manual input from 0.0 to 50.0 ppt; automatic correction
|
|
Barometric Pressure Correction
|
manual input from 0.0 to 50.0 ppt; automatic correction
|
|
Temperature Sensor
|
100 Pt RTD
|
|
Temperature Compensation
|
Automatic or manual from -9.9 to 125°C
|
|
Password Protection
|
two level password
|
|
Display
|
dual line LCD, 4-digit upper and 3 1/2 digit lower
|
|
Operating Temperature
|
-10 to 60°C
|
|
Housing
|
1/4 DIN
|
|
Dimensions
|
3.8" x 3.8" x 6.9" (9.6 x 9.6 x 17.5 cm)
|
|
Shipping Weight
|
2.5 lbs (1.2 kg)
|